Year : 2007 | Volume
: 50
| Issue : 3 | Page : 634-5 |
|
Validation of ESR analyzer using Westergren ESR method.
M Sikka, R Tandon, U Rusia, N Madan
Department of Pathology, University College of Medical Sciences and GTB Hospital, Delhi
Correspondence Address:
M Sikka Department of Pathology, University College of Medical Sciences and GTB Hospital, Delhi
 Source of Support: None, Conflict of Interest: None  | Check |
PMID: 17883168 
|
|
Erythrocyte sedimentation rate (ESR) is one of the most frequently ordered laboratory test. ESR analyzers were developed to provide a quick and efficient measure of ESR. We compared the results of ESR obtained by an ESR analyzer with those by the Westergren method in a group of 75 patients Linear regression analysis showed a good correlation between the two results (r = 0.818, p < 0.01). The intra class correlation was 0.82. The analyzer method had the advantages of safety, decreased technician time and improved patient care by providing quick results. |
|
|
|
[PDF Not available]* |
|
 |
|